5 – 25Operation Manual KjelDigester K-446/K-449
5 Installation
5 Installation
5.1 Unpacking
DANGER
Risk pacemaker malfunction!
Death or serious injuries can result.
1. Keep a distance of at least 15cm between a pacemaker and the magnets of
the rack support.
CAUTION
Risk of injuries due to weight of the device!
Minor or moderate injuries can result.
1. Call a second person to lift the device.
2. Do not lift the device alone.
1. Unpack the device.
CAUTION
Risk of cuts from damaged glass parts!
Minor or moderate cuts can result.
1. Handle glass parts with care.
2. Check all glass parts for damage before installation.
3. Do not use any damaged parts.
4. Do not touch cracks or bits of broken glass with bare hands.
2. Check the device for any damage caused during the transportation.
3. Report damage to the responsible transportation company.
4. Keep the original packaging for future transportation.
5. Check all glass parts for damage.
8.4.2 Glass parts, check, (858).
5.2 Preparation for assembly
1. Place the device in a fume hood.
It is recommended that the fume hood complies with the norm DINEN14175-2.
Keep around the device a free space of at least 5cm.
2. If using a Scrubber, place it on the left side of the KjelDigester.
NOTE
Keep the Scrubber suction hose as short and straight as possible. It prevents cooling air of
the Scrubber blowing towards the heating block of the KjelDigester.
3. Remove the transport lock, (5–26).