BÜCHILabortechnikAG Installation|5
PrepChromC700 49/110
u MovetheupperholderdownontothecolumnandtightenthePEEK
adapter(5).
u Tightentheupperknurledknob(1).
5.6 Mixingchambervolume,adjustment
NOTE
Theadjustmentofthemixingchamberisrecommendedtominimizethedead
volume.
u Turnthewheel(2)toadjustthemixingchamber(3)accordingtothede
siredflowrateofthepump.
Turningthewheelclockwisewillincreasethemixingvalue.
Turningthewheelcounterclockwisewilldecreasethemixingvalue.
u Readthecurrentvalueonthescaleoftheprotrudingpiston(1).
NOTE
Thepistoncanbeturnedwitha11/32"spannerifthescaleisturnedtowards
thedevice.
Flowrate Mixingchamber
0–80mL/min 0%
80–130mL/min 40%
130–175mL/min 80%
>175mL/min 100%