BÜCHILabortechnikAG Helpwithfaults|8
OperationManualRotavapor®R100 43/60
Fault Possiblecause Remedy
Optimumdistilla
tionnot
achieved
Toomuchbackevaporation
fromthereceivingflask(espe
ciallywithsolventmixtures)
u Emptyreceivingflaskand
restartdistillation(seeChap
ter6.3.2"Removingthere
ceivingflask",page38,
Chapter6.1.4"Fittingthere
ceivingflask",page30and
Chapter6.2"Performingdis
tillation",page32).
Otherfaultsindistillation
process(e.g.suddencooling,
toolittleheatflow,etc.)
u Checktemperaturesofheat
ingbathandcoolantandad
justasnecessary(see
Chapter6.1.1"Preparingthe
heatingbath",page27and/
orcooleroperatinginstruc
tions).
u Lowerpressureuntildistilla
tionresumes(seevacuum
pumpoperatinginstructions).