7 Maintenance and repairs
46 K-425 / K-436 Operation Manual, Version A
7.5.2 500 ml sample tube support spring
The 500 ml tube rack is equipped with a tube holder containing five sample tube support springs.
Each flat spring must be freely rotatable by a small amount. If swivelling is not possible, exchange the
rake!
Note
Rinse the tube holder with water after use to avoid corrosion.•
Store it dry and clean after use.•
7.6 Protection cover at the front
The protection cover at the front is equipped with double-faced adhesive tape. If it is strongly contami-
nated or damaged, replace it with a new one.