5|Installation BÜCHILabortechnikAG
18/42 OperationManualInterfaceI100
5.2 Connections
1
1
2
3
4
5
1
6
1
7
Fig.9:ConnectionsonI100
u Connectthecommunicationcablefromthevacuumpumptotheconnection
markedAS/SB(1).
u ConnecttheleadfromthecoolantwatervalvetotheRS485socket(6)ontherear
panel,ifused.
u Connectthecommunicationcablefromthevalveunittotheconnectionmarked
VALVE(3),ifused.
u Connecttheleadfromthepoweradaptertothe30VDCsocket(4).
u Connectthevacuumtubingforpressuremeasurementtothevacuumconnection
(5)withGL14capnut.
u Ifinertgasisrequired,connectittohosebarb(7).
5.3 Connectiontovacuumpump
ForconnectingtheInterfaceI100totheVacuumPumpV100werecommendusinga
Woulffbottle.